chuin@redmarkindustry.com

8 channel Pipette

Previous 9 / 15 Next

8 channel Pipette  Others


Fully Autoclave:
The MicroPette plus pipettes can be fully autoclaved at 121¡æ, 1 atm for minutes. The pipettes can be autoclaved without special preparations. After autoclaving the pipette must be cooled down and dried over 12 hours.

Features for Multi-channel pipette
  •  8 and 12 channel pipettes are available for standard 96-well plate 
  •  Individual piston and tip cone assemblies allow easy repair and maintenance 
  •  Compound material tip cone design allows visual seal vertification
  •  Universal pipette tips can be used 
  •  Dispensing head rotates ensure optimum pipetting convenience 
Technical Parameters: 
8-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume
(μl)
Maximume Permissible Systematic error
  (Inaccuracy)
Maximume Permissible Random error
(Inaccuracy)
      % μl % μl
0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4
12-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume
(μl)
Maximume Permissible Systematic error
   (Inaccuracy)
Maximume Permissible Random error
(Inaccuracy)
      % μl % μl
0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4
Fixed Volume Pipettes
Volume Range Increment Test Volume
(μl)
Maximume Permissible Systematic error
   (Inaccuracy)
Maximume Permissible Random error
(Inaccuracy)
      % μl % μl
5μl - 5μl 1.3% 0.065 1.2% 0.06
10μl - 10μl 0.8% 0.08 0.8% 0.08
20μl - 20μl 0.6% 0.12 0.5% 0.1
25μl - 25μl 0.5% 0.125 0.3% 0.075
50μl - 50μl 0.5% 0.25 0.3% 0.15
100μl - 100μl 0.5% 0.5 0.3% 0.3
200μl - 200μl 0.4% 0.8 0.2% 0.4
250μl -- 250μl 0.4% 1.0 0.2% 0.5
500μl - 500μl 0.3% 1.5 0.2% 1.0
1000μl - 1000μl 0.3% 3.0 0.2% 2.0
2000μl - 2000μl 0.3% 6.0 0.15% 3.0
5000μl - 5000μl 0.3% 15 0.15% 7.5
 